SINGLE & DUAL RANGE
DIALYSATE METERS

The choice of hemodialysis professionals for years, these compact instruments have been designed specifically to test dialysate solutions. By measuring electrical conductivity, they will quickly determine if dialysate concentrations are correct.

To use these hand-held instruments, the operator simply fills the cell cup with dialysate solution and pushes a button. The needle indicates the total concentrations of salts in the solution.

Features:

Fast and accurate dialysate  tests

Models for both acetate and bicarbonate  solutions

Chip/crack resistant polyethylene cell  cup

Rugged design

Simple to use

Automatic temperature  compensation

Accessories

PTP

Porta-Pak (ABS Plastic  Carrying Case)

NaCl 12.5, 13.4 or 14.0 mS  Standard Solution 1 quart bottles

NaCl 12.5, 13.4 or 14.0 mS  Standard Solution 1 gallon bottles

Solutions

Solutions

Standard solutions are precisely formulated and are used to recalibrate Dialysate  Meters for greatest accuracy.  The 12.5, 13.4 or 14.0 millimhos Standards are  the same used in Myron L factory calibration of Models D-1 and  D-2.

Dialysate Meter

Has an expanded scale of 11-15 millimhos  and is useful for testing final dialysate.  A unique feature of this model is  the adjustable green color band above the dial.  This provides an immediate  visual confirmation of whether or not the dialysate concentration is within the  patient's "safe zone".

Model D-1

Buy Now

The choice of hemodialysis professionals for years, these compact instruments have been designed specifically to test dialysate solutions.  By measuring electrical conductivity, they will quickly determine if dialysate concentrations are correct.
Model D-2 

Features dual scales on the dial and can test both acetate and  bicarbonate dialysate.  Both instruments are easy to calibrate for more accurate  readings.

Both instruments have built-in cell cups  and are completely self-contained. They are powered by 9 volt batteries which  last for at least 2,000 tests.  A glow light serves as a battery reserve  indicator.

Thousands of MYRON L Dialysate Meters are  currently in daily use around the world after years of reliable service.

Instrument  Specifications

Ranges

Model D-1

11-15 millimhos

Model D-2

0-10 millimhos

0-20  millimhos

Accuracy

± 2% of full scale
± 1% at 14.0 millimhos calibration  point

Repeatability

±1%

Temperature Compensation

Automatic (to 25ºC / 77ºF) between 50º and 160ºF  (10-71ºC)

Electrodes

Built-in, never need replatinizing

Calibration

Easy & fast with 14.0 millimhos Standard Solution and  finger adjustable

 internal trimmer (also 12.5 and 13.4 millimhos solution  available)

Circuitry

Solid-state, sealed against moisture

Power

Two 9 volt batteries supplied

Battery Life

Approximately one year or 2,000 tests

Case

Heavy gauge, textured ABS

Cell Cup

Chip and crack resistant polyethylene

Dimensions

3.4" W x 4.5" D x 4.0" H (85 x 129 x 126 mm)

Weight

One pound (450  gms)

Limited Warranty

Myron L Dialysate Meters have  a two year warranty period.  If an instrument fails to function normally, return  to factory prepaid.  If, in the opinion of the factory, failure was due to  materials or workmanship, repair or replacement will be made without charge.  A  reasonable service charge will be made for diagnosis or repairs due to normal  wear, abuse, or tampering.  Warranty is limited to the repair or replacement of  the Myron L Dialysate Meter only.  The Myron L Company assumes no other  responsibility or liability.